| MEDICAL EQUIPMENT FAILURES |
|
MEDICAL EQUIPMENT FAILURE
|
CLINICAL RESPONSE
|
OTHER ACTIONS
|
| Defibrillator |
Resume CPR and ACLS measures until alternative defibrillator is brought to the bedside. |
Every department should know where the closest “BACK-UP” is located (i.e., down the hall, next room, etc.) |
| Monitor |
1) Retrieve a transport monitor and connect the
patient
OR
2) Retrieve defibrillator. The monitor portion can be connected to patient for monitoring purposes. |
After Clinical Engineering hours, notify the Nursing Supervisor. Clinical
Engineering will either respond by phone or Nursing Supervisor will know the location of Clinical Engineering’s spare roll-around monitor. |
| Ventilator |
Hand-ventilate patient until spare ventilator arrives. |
Notify the Respirator Therapist assigned to the specific unit or page the Therapist-in-Charge (719-4946) to obtain spare ventilator. |
| Dialysis Machine |
Remove blood-line from venous line clamp. Then manually hand-crank blood-pump to return patient’s blood. |
After hours, obtain another dialysis machine. The dialysis technician will investigate the malfunctioning unit the following work day. |
| Heart-Lung Machine and/or ECMO Pump |
Electrical Failure
Utilize hand crank.
Mechanical Failure (crank does not work)
Remove tubing from the malfunctioning pump head and place in other pump. |
Heart-Lung Machines” Replace malfunctioning pump with spare pump (located in pump room). Notify Clinical Engineering OR Shop (353-1736)
ECMO: Change out malfunctioning equipment with spare. Contact Clinical Engineering main shop. (353-1727) |
| Anesthesia Unit |
Ventilator
Hand-bag patient until spare unit arrives.
Gas Wall-Outlet Failure
Connect patient to the E-tank (located on cart). |
Obtain spare unit. Notify Clinical Engineering (Parnassus
353-1736) or (Mount Zion 885-7642).
Notify
Medical Center Facilities (Parnassus –3-1120) or
(Mount Zion – 5- 7576). |
| Ventricular Assist Device (VAD) |
Thoratec
Utilize hand bulbs in back of machine until spare arrives.
BioMedicus Utilize hand crank (2 cranks) until spare arrives. |
The Medical Center has a total of three
(3) machines (2 Thoratecs and 1 Bio-Medicus), all located in the Operating
Room. Notify Clinical Engineering OR Shop (353-1736). |
| Continuous Veno/Venous Hemofiltration (CVVH) |
Utilize hand crank to prevent blood from clotting until spare pump arrives |
Contact Central Services 353-1837 to obtain a replacement pump. |
Continuous Veno/Venous Hemofiltration (CVVH) |
Obtain replacement pump immediately. |
The Medical Center has three (3) IAPBs (two located in Intensive Cardiac Care and one in the Operating Room). |
| Emergency Aspirator |
Hand bag patient until spare unit arrives. |
Contact Central Service (Parnassus 353-1837) or (Mount Zion 885-7255) to request an emergency replacement aspirator. |
| Pumps, Infusion or PCA |
Stop pump. |
Contact Central Services (Parnassus 353-1837) or Mount Zion (885-7255) to request an emergency replacement pump. |
| ALL other equipment |
Stop using equipment and attain spare equipment from appropriate source. |
If after hours and device needs immediate repair, notify Nursing Supervisor in order to get a Clinical Engineer to respond to problem. |
| These are general guidelines
and should be reviewed for clinical practices that are specific to
your department. |
|